Introduction
This is a compact peristaltic pump designed for laboratory applications, featuring a rotary knob for speed adjustment within a range of 0.1–350 rpm. It is compatible with KT15 and series pump heads. Housed in a durable ABS enclosure, the pump supports bidirectional fluid transfer for flexible operation. By replacing the pump head tubing with different sizes, it can achieve a maximum flow rate of 326.55 mL/min, delivering stable and precise liquid dispensing performance. Ideal for scenarios requiring consistent dripping in laboratory settings.
Applications
Features
- KT15 Head: Flip-top design for easy opening and fast tube replacement.
- Flow Range: Up to 326.55 mL/min, mainly determined by the installed tubing size.
- LED Screen: Displays speed. Easy rotary knob speed adjustment (0.1-350 RPM).
- ABS Engineering Plastic Enclosure: Corrosion-resistant and anti-static.
- Communication: USB & RS485 (Modbus RTU protocol).
- Key Features: Speed adjustment, start/stop control, and direction reversal.
Flow Range
| Drive Model | Pump Head Model | Speed range (rpm) | ID × Wall Thickness (mm) | Flow Rate Range (ml/min) |
| LabT | KT15 | 0.1-350 | 13# | 0.0033-11.55 |
| 14# | 0.0156-54.60 | |||
| 19# | 0.0286-100.10 | |||
| 16# | 0.0477-166.95 | |||
| 25# | 0.0933-326.55 |
*To view the full table on mobile, please swipe left or right on the screen.
*The purchase of this peristaltic pump includes five complimentary tubing sections, each 2 meters long. The included sizes are: #13, #14, #16, #19, and #25.
Technical Data
| Display | OLED | Control | Knob+Button |
| Communication | USB, RS485 (Modbus RTU protocol) | Condition Temperature | 0-40℃ |
| Housing Material | ABS | IP Rate | IP31 |
| Speed Range | 0.1-350rpm, adjustable | Power Supply | AC100V-240V |
| Speed Resolution | 0.1rpm | Power Consumption | ≤30W |
| External Speed Control Signal | 0-5V, 4-20mA, 0-10V(optional) | Gross Weight | 3.31 lb |
*To view the full table on mobile, please swipe left or right on the screen.
Precautions
1.Screen does not light up when the switch is turned on
Check if the power cable is properly and securely connected./Confirm that there is power supply./Check if the fuse is damaged.
2.Pump head does not rotate when the start button is pressed
Check if the pump head screws are tightened./Check for any foreign objects blocking the pump head rollers./Check if the roller bearings are severely corroded and unable to rotate (check after power off)./Confirm if the machine is in external control mode (refer to external control instructions).
3.No fluid transfer during operation
Check if the hose installed is compatible with the pump head./Confirm if there is enough fluid at the inlet (fluid level above the pipe inlet)./Check if there is any damage or leakage in the pipe and replace or seal it./Check if any solid material is blocking the pipe.
4.Actual fluid output does not match the preset value
Check if the hose settings for the pump head match the actual installation and usage./If using external control, confirm if the external control signal is complete or has attenuation.