Introduction
The U.S. Solid USS-DBS62-1 Analytical Balance delivers high precision with 0.0001g accuracy and a 100g capacity. Constructed with durable materials, it ensures quick and efficient sample weighing. Featuring electromagnetic force compensation technology, it provides reliable measurements for diverse lab applications. Equipped with RS-232 and USB ports for data transfer, it supports functions like percent weighing, density weighing, underhook weighing, and dynamic weighing to meet various laboratory needs.
Applications


Features
- Readability: 0.0001g with electromagnetic force compensation technology
- Capacity:0.1mg-100g
- Stabilization Time: ≤2 seconds
- External Calibration: Manual calibration with weights
- 8 Weighing Units Conversion: g, oz, ct, mg, lb, gn, ozt, dwt
- RS232 and USB
- Built-in Applications: piece counting, percent weighing, animal/dynamic weighing, density determination, sensitivity setting and etc.
- Underhook Weighing: Easily determine the mass of liquids, unstable samples, and other challenging substances—ideal for density calculations. (Includes a built-in hook for attaching custom hanging tools; sample hooks not included.)
- Powered by AC (100V-240V)
More Details


User Manual
To learn how to install, calibrate, set up, and use this product, please refer to the user manual.
Packing List
① | Balance Main Body | ② | Power Cable | ||
③ | Calibration Weight | ④ | Instruction Manual | ||
⑤ | Weighing Pan |
*To view the full table on mobile, please swipe left or right on the screen.
Technical Data
Model | USS-DBS62-1 | Working Temperature | 63.5°F - 72.5°F | ||
Weighing Sensor | Electromagnetic Force | Stabilization Time | ≤2s | ||
Calibration | External calibration | Calibration Weight | Yes | ||
Capacity | 0.1mg-100 g | Units | g, oz, ct, mg, lb, gn, ozt, dwt | ||
Readability | 0.0001g | Interface | RS232 and USB | ||
Repeatability | ±0.1mg | Power | AC100-240V 50/60Hz | ||
Linearity | ±0.2mg | Packaging Dimensions | 16.54"x14.57"x16.73" | ||
Pan Size | Φ90 mm | Gross Weight | 17.64 lbs | ||
Number of Displays | 1 | Windshield | Yes |
*To view the full table on mobile, please swipe left or right on the screen.
Precautions
- Leveling the Balance: Accurate horizontal positioning necessary for repeatable measurementsand exact results. Adjust the leveling feet of the analytical balance until the air bubbleinthe indicator is centered.
- Warm-Up Requirement: For accurate results, the balance must warm up for at least 30 minutes after being powered on or after a power outage longer than 30 minutes, to reach the proper operating temperature before calibration or use.(For warm-up time details, refer to the user manual or contact support.)
- Stability: Place the balance on a stable, vibration-free surface, away from drafts and electromagnetic interference. It's important to note that high-precision balances are more susceptible to environmental factors, which can affect reading stability.
- Calibration: Calibrate the balance if it is moved (even within the same room), when ambient conditions change, and before each use. If pressing the "CAL" button has no effect, ensure you are pressing and holding it. For most models, calibration requires a long press rather than a quick tap.(Refer to the manual for specific instructions).
Troubleshooting
Problem | Cause | Solution | |||
Display not working | No power supply | Plug in the adapter | |||
Fuse damaged | Replace the fuse | ||||
Damaged transformer | Replace the power transformer. lf problem persists, send to Technical Services for repair. | ||||
Unstable Display | Poor working conditions | Seek somewhere with less vibration and/or airflow | |||
Wind or drafts affecting readings | Check to make sure draft shield is closed tightly | ||||
Scale pan unsteady | Remove the pan and clean the surface of the balance well, making sure there are no lingering items | ||||
Power exceeds permissible value | Connect balance to power supply AC110-220V | ||||
Displayed mass is not the same as true mass | Balance has not been calibrated | Double check calibration, internal and/or external | |||
Receptacle not being considered | Make sure the receptacle has been tared out | ||||
Uneven surface | Make sure balance is on a flat, even surface |
*To view the full table on mobile, please swipe left or right on the screen.
Shipping Time
The processing and delivery time of this item is about 20 working days.
All Options
Units | Model | Readability | Capacity | Repeatability | Linearity | Stabilization Time | Calibration Weight | Calibration | Interface | Pan Size | Power Supply |
8 types | USS-DBS62-1 | 0.0001g | 0.1mg-100g | ±0.1mg | ±0.2mg | ≤2s | yes | External calibration | RS232 and USB | Φ90mm | AC100-240V |
USS-DBS62-1.2 | 0.1mg-120g | ||||||||||
USS-DBS62-2 | 0.1mg-200g | ||||||||||
USS-DBS62-2.2 | 0.1mg-220g | ||||||||||
USS-DBS62-3 | 0.1mg-300g | ||||||||||
USS-DBS62-3.2 | 0.1mg-320g | ±0.2mg | ±0.3mg | ||||||||
USS-DBS62-4 | 0.1mg-400g | ||||||||||
USS-DBS62-4.2 | 0.1mg-420g | ||||||||||
USS-DBS62-5 | 0.1mg-500g | ||||||||||
USS-DBS62-5.2 | 0.1mg-520g | ||||||||||
USS-DBS63-6 | 0.001g | 1mg-600g | ±1mg | ±2mg | ≤1s | no | |||||
USS-DBS63-8 | 1mg-800g | ||||||||||
USS-DBS63-10 | 1mg-1kg | ||||||||||
USS-DBS63-20 | 1mg-2kg | ±3mg | ±4mg | ≤3s | |||||||
* 8 Weighing Units Conversion: g/oz/ct/mg/lb/gn/ozt/dwt |